Analysis of PPM Levels of Hydrogen in Liquid Mixtures of Propylene and/or Ethylene APPLICATION # 130
The Analysis This GC is used to analyze approximately 5 ppm (v/v) of hydrogen in liquid propylene mixtures.
How It
Works The GC is configured to receive liquid injections directly from the pressurized Welker cylinder. A valve is actuated and it injects the sample, which passes through another valve and into a molecular sieve column to the TCD detector. The mole-sieve column is protected with a pre-column in order to avoid the matrix hydrocarbon from entering this column. Total analysis time is approximately 8 minutes.
Other Uses1. Separate H2 from ethylene and acetylene gas mixtures. 2. Analyze H2 at 1.0-1000 ppm in gas mixtures of ethylene and acetylene. 3. Analyze H2 at 1.0 ppm (by weight) in liquid propylene.
|
|
|